Most X-ray scanner failures are preventable and start with three things: a bent lead curtain, a blocked emergency stop, and a dirty detector. A five-minute check at the start of each shift catches all three, and it costs less than one hour of downtime.
Monthly: check the radiation warning lamp, verify the door interlock, inspect cables and rollers. Annually: a third-party radiation survey, a full image-quality verification with the test piece, and a review of the spare-parts stock (curtains, belts, fans).
Pushing bags through the curtain. Forcing an oversized bag bends the strips and opens a gap. Train staff to lift the curtain, not push through it.
Ignoring a drifting belt. A belt that tracks to one side wears the edge and eventually jams under load — usually at the worst possible moment.
Cleaning with the wrong cloth. Abrasive or solvent cleaning fogs the detector cover and permanently degrades image quality.
How often should curtains be replaced? When a strip no longer lies flat, or at least once a year on busy lanes. Keep a spare set on site.
Do operators need dosimeters? Recommended for anyone working beside the machine daily; a wearable dosimeter gives a record if a question is ever raised.
What if the image looks washed out? Check curtain condition and detector cover first, then run the test piece. If the test piece is also washed out, call service — do not compensate by turning up the gain.