CZT stands for cadmium zinc telluride (CdZnTe), a compound semiconductor that detects X-rays and gamma rays at room temperature. Conventional detectors use a scintillator to convert radiation into visible light and then a photomultiplier or photodiode to read it. CZT converts radiation directly into charge, removing the light-conversion step, which makes the detector smaller, more energy-resolving and able to work at lower dose.
| Generation | Technology | Form | Limitation |
|---|---|---|---|
| First | Gas detectors (ion chamber, proportional counter) | Bulky, high voltage | Low efficiency, poor for imaging |
| Second | Scintillator + photomultiplier | Heavy, shock sensitive | Large, high power |
| Third | Scintillator + photodiode | Compact, array capable | Limited energy resolution |
| Fourth | Semiconductor (CZT) | Thumb-sized module | Large crystals remain costly |
| Model | Use | Dose rate range | Energy range | Size |
|---|---|---|---|---|
| DR-02B | General dose monitoring | 0.1 μSv/h – 10 mSv/h | 30 keV – 3 MeV | 29.16 × 25 × 3.06 mm |
| DR-04A | Alpha / radon monitoring | — | Alpha particles | 29 × 28 × 2.3 mm |
| DR-05A/B | Energy-compensated dose | 0.1 μSv/h – 100 mSv/h | 30 keV – 3 MeV | 27.56 × 25 × 3 mm |
| GZHY-SM-01A | General module | 0.01 μSv/h – 20 mSv/h | 30 keV – 3 MeV | 37.96 × 17.10 × 3.50 mm |
| GZHY-CSPre-01A | Charge-sensitive preamplifier | — | CZT and Si detectors | 28 × 16 × 6.6 mm |
Modules run from 3.3–5 V DC and communicate over industrial serial UART, connecting through an FPC cable for easy integration.
| Item | CZT | Scintillator + photodiode |
|---|---|---|
| Energy resolution | Good; spectroscopy and multi-energy imaging | Moderate; counting and dose only |
| Size | Small, fits wearables | Larger |
| Cost | Higher for large crystals | Lower for large arrays |
| Best fit | Spectroscopy, dosimetry, portables | Large-area imaging on a budget |
FjXray works with NPU (Photonuclear Detector) to supply these CZT modules and complete devices, including custom development and algorithm support. When selecting a model, tell us what you measure (X-ray, gamma or alpha), the dose-rate range and the available space.
Send us the application, measurement range and device format, and we will propose the module and integration approach.